品牌
其他厂商性质
深圳市所在地
Measuring principle | ZrO2 sensor with Pt-electrodes
|
Measuring range | 0-25%Vol |
Analogue signal port | 0 … 10 V, 0 (4) …20mA, galvanically separated
|
Reproducability | +/- 2 % of the measuring value
|
Resolution
| 0,01 % – C(O2) |
T90-Time
| ca. 20 seconds |
Display | 2* 16 digit, illuminated LCD display 1. Line: display of concentration in Vol % 2. Line: messages, device status, sensor parameters |
Messages | 1 System message (measuring value yes / no) max. 2 messages configurable as measuring values |
Digital communication | serial interface RS232 |
ambient operating temp.
| +5 °C up to + 45 °C |
Power supply
| 110 or 230 VAC / 60 - 50 Hz |
Protection / Housing / Dimensions | IP 65/ wall mounting housing 300x320 x130mm IP20 / electronics unit 28 TE / 3HU IP54 / portable housing 42 TE / 3 HE IP 54 / panel mounting 144 x 144 mm
|
Weight
| 2.5-10 kg |
Options
| auto-calibration back purge for high dust concentrations pressure reducer particle lfilter 2-7μm manual 5 way valve electronical/pneumatic gas supply pump electronical flowmeter software for remote control and service
|
Approvals TÜV certification 936/807023/A of 14.08.2000
Version: AMS 3220 E V-2013-07
Specification subject to change.
The Measuring principle:
The Zircondioxid sensor is the ideal detector for the measurement of Oxygen in gases without
combustible impurities. The Zircondioxid sensor can be installed in the housing of the Oxygen
Analyser AMS 3220 . Using a calibration adapter with the Zircondioxid sensor, the Zircondioxid sensor can also be installed directly in a process gas pipe. This setup allows easy supply of calibration and reference gas to the sensor by using a separate pneumatic unit. In combination with the optional, time controlled auto-calibration feature the already high accuracy and reproducibility of the Oxygen Analyser AMS 3220 is improved further.
The Measuring system:
A long list of options allows to adapt the Oxygen Analyser AMS 3220 for almost every application.
Four measuring ranges with automatic switch-over and digital measuring range recognition can be set freely within the maximum measuring range. The software for remote control and remote service with access via the customer computer network further increase the flexibility of the Oxygen Analyser AMS 3220 in a application.