Multi-channel Analyzer | 12bit, 80Msps AD. 4096 channels analyze ability |
with FPGA hardware and robust DSP algorithm | |
to discriminate signal and noise | |
Maximum count rate | FPC:2Mcps SPC:1Mcps SC: 1.5Mcps |
Pulse shift & Gain correction | Automatic |
Dead time correction | Automatic |
Sample Chamber | X-ray tubeMg | ||||
Type | Vacuum for solids | Windows | The thin beryllium window, 50um | ||
Dimensions | 51.5mm Øx40mm height, max. | or 75um, ultra-high transmission | |||
Weight | Max. 500g including sample holder | Rhodium (Rh) as standard, option | |||
Dual position with anti-dust filter & | |||||
Anode target material | includes Copper (Cu), | ||||
Sample loading | pre-vacuum pumping as standard | Molybdenum (Mo), Tungsten (W), | |||
feature. Analyzing one sample | Chromium (Cr), Platinum (Pt) | ||||
while pumping the next one | |||||
Sample changer | Automation robotic changer system, | Operation | Tube remains powered on during | ||
high capacity up to 168 samples | sample holding | ||||
3 spinning speed modes | Dual tube cooling water circulation, | ||||
Spinner | Cooling water | conductivity of deionized water is | |||
(low, medium, high) | |||||
much lower than 1uS | |||||
HV Generator | Rating | ||||
Output | Selectable in steps of 1KV, 1mA | Power | |||
as standard. Fine step is optional | 4kW | ||||
by 12bit DA | |||||
Long Term Stability | 0.01%/8 hours | KV/mA | 20-60KV, 75kV is optional | ||
Temperature Coefficient | 50 ppm/°C (20 ppm/°C optional) | 10-140mA, 150mA is optional | |||
Goniometer | Optical Path | ||||
Innovative steel-belt-drive system, | Channels masks | Single mask | |||
Type | no friction, no backlash, | (fixed 27, 30 or 37mm) | |||
feedback by optical encoder, | Primary collimators | 3 max: 100, 150, 300, 550, | |||
0/20 independent | 700 or 4000 um, selectable | ||||
Maximum slewing speed | 6000 • 20/min | Primary beam filters | 4 max: Pb, Al, Cu with different | ||
thickness | |||||
Angular accuracy | Crystals | 8 max: LiF420, LiF220, LiF200, | |||
0.003° 0 and 20 | Ge111, PE002, lnSb, TIAP, | ||||
multilayers for light element | |||||
Angular reproducibility | 0.0002° 0 and 20 | 3 max: Flow proportional | |||
Detector | counter (FPC), Sealed Xe | ||||
Step scan range | Min. 0.0001°, Max.1° | proportional counter (SPC), | |||
Scintillation counter (SC) | |||||
Scanning angle range | PC:17° to 146°20 | Optical Cabinet Past | Vacuum < 10Pa | ||
SC: 0° to 120°20 |