品牌
其他厂商性质
上海所在地
Circuit | Custom one-chip of microprocessor LSI circuit. | |||
Display | LCD size : 52 mm x 38 mm LCD with green backlight ( ON/OFF ). | |||
Measurement | Velocity, Acceleration, Displacement | |||
Function | Acceleration, Velocity : RMS, Peak, Max. Hold. Displacement : p-p ( peak-peak ), Max-hold p-p. | |||
Unit | Measurement | Metric | Imperial | |
Acceleration | meter/s^2,G | ft/s^2, | ||
Velocity | mm/s, cm/s | inch/s | ||
Displacement | mm | inch | ||
Frequency range | 10 Hz to 1 KHz * Sensitivity relative during the the frequency range meet ISO 2954 Refer to table 1, page . | |||
Circuit | Exclusive microcomputer circuit. | |||
Peak Measurement | Acceleration, Velocity : To measure and update the peak value. Displacement : To measure and update the peak to peak ( p-p ) value. | |||
Max. Hold Measurement | Acceleration, Velocity : To measure and update the max. peak value. | |||
Displacement : To measure and update the max. peak to peak ( p-p ) value. | ||||
Zero Button, Max. Hold Rest Button | Build in the Zero button and the Max. Hold reset button on front panel. | |||
Datalogger Sampling Time Setting range | Auto | 1 second to 3600 seconds @ Sampling time can set to 1 second, but memory data may loss. | ||
Manual | Push the data logger button once will save data one time. @ Set the sampling time to 0 second. @ Manual mode, can also select the 1 to 99 position ( Location ) no. | |||
Memory Card | SD memory card. 1 GB to 16 GB. | |||
Advanced setting | Set clock time ( Year/Month/Date, Hour/Minute/ Second ) Decimal point of SD card setting Auto power OFF management Set beep Sound ON/OFF Set sampling time SD memory card Format | |||
Data Hold | Freeze the display reading. | |||
Memory Recall | Maximum & Minimum value. | |||
Sampling Time of Display | Approx. 1 second. | |||
Data Output | RS 232/USB PC computer interface. Connect the optional RS232 cable UPCB-02 will get the RS232 plug. Connect the optional USB cable USB-01 will get the USB plug. | |||
Operating Temperature | 0 to 50 °C. | |||
Operating Humidity | Less than 85% R.H. | |||
Power Supply | * Alkaline or heavy duty DC 1.5 V battery ( UM3, AA ) x 6 PCs, or equivalent. | |||
* DC 9V adapter input. ( AC/DC power adapter is optional ). |
Power Current | Normal operation ( w/o SD card save data and LCD Backlight is OFF) : Approx. DC 15 mA. |
When SD card save the data and LCD Backlight is OFF) : Approx. DC 36 mA. | |
Weight | Meter : 515 g/ 1.13 LB. |
Probe with cable and magnetic base : 99 g/0,22 LB | |
Dimension | Meter : 203 x 76 x 38 mm |
Vibration sensor probe: Round 16 mm Dia. x 37 mm. Cable length : 1.2 meter. |
Unit | m/s^2 |
Range | 0.5 to 199.9 m/s^2 |
Resolution | 0.1 m/s^2 |
Accuracy | ± ( 5 % + 2 d ) reading @ 160 Hz, 80 Hz, 23 ± 5 °C |
Calibration Point | 50 m/S^2 ( 160 Hz ) |
Unit | G @ 1 G = 9.8 m/s^2 |
Range | 0.05 to 20.39 G |
Resolution | 0.01 G |
Accuracy | ± ( 5 % + 2 d ) reading @ 160 Hz, 80 Hz, 23 ± 5 °C |
Calibration Point | 50 m/S^2 ( 160 Hz ) |
Unit | ft/s^2 |
Range | 2 to 656 ft/s^2 |
Resolution | 1 ft/s^2 |
Accuracy | ± ( 5 % + 2 d ) reading @ 160 Hz, 80 Hz, 23 ± 5 °C |
Calibration Point | 50 m/S^2 ( 160 Hz ) |
Remark : RMS : To measure the true RMS value. Peak : To measure and update the peak value. Max. Hold : To measure and update the max. peak value. |
Unit | mm/s |
Range | 0.5 to 199.9 mm/s |
Resolution | 0. 1 mm/s |
Accuracy | ± ( 5 % + 2 d ) reading @ 160 Hz, 80 Hz, 23 ± 5 °C |
Calibration Point | 50 mm/s ( 160 Hz ) |
Unit | cm/s |
Range | 0.05 to 19.99 cm/s |
Resolution | 0. 01 cm/s |
Accuracy | ± ( 5 % + 2 d ) reading @ 160 Hz, 80 Hz, 23 ± 5 °C |
Calibration Point | 50 mm/s ( 160 Hz ) |
Unit | inch/s |
Range | 0.02 to 7.87 inch/s |
Resolution | 0.01 inch/s |
Accuracy | ± ( 5 % + 2 d ) reading @ 160 Hz, 80 Hz, 23 ± 5 °C |
Calibration Point | 50 mm/s ( 160 Hz ) |
Remark : RMS : To measure the true RMS value. Peak : To measure and update the peak value. Max. Hold : To measure and update the max. peak value. |